MicronTracker 3 Specifications

Below you can find the detailed specifications for the available MicronTracker models. The highlighted rows are some details that are generally more important to consider. Please contact us for more information and if you need any advice on choosing the model to best suit your specific application.
Specification \ Model H3-60 Sx60 Hx40 Hx60
Available now now now now
FOM, spherical section (radius x width x height) 240 x 200 x 160 cm 115 x 70 x 55 cm 120 x 120 x 90 cm 200 x 130 x 100 cm
Measurement rate 16 Hz 48 Hz 20 Hz 20 Hz
Calibration accuracy1 0.20 mm RMS 0.25 mm RMS 0.20 mm RMS 0.35 mm RMS
Jitter, static target2 0.007 mm RMS 0.007 mm RMS 0.015 mm RMS 0.015 mm RMS
Jitter, moving target2 0.07 mm RMS 0.07 mm RMS 0.14 mm RMS 0.14 mm RMS
Processing time/frame (2.5GHz Intel Q8300, 4 markers) 15 ms 7 ms 10 ms 10 ms
Lag ~60 ms ~20 ms ~50 ms ~50 ms
Sensors resolution 1280×960 640×480 1024×768 1024×768
Lenses / H°xV° 6 mm 50°x38° 6 mm 50°x38° 4 mm 70°x52° 6 mm 50°x38°
Case dimensions (WxHxD, approx.) 283 x 43 x 49 mm 164 x 43 x 54 mm 164 x 43 x 54 mm 164 x 43 x 54 mm
Weight (approx.) 505g 342g 342g 342g
Electrical Interface IEEE-1394b (FireWire), 800 Mbps IEEE-1394a (FireWire), 400 Mbps
Mount 1/4″ thread tripod mount
Operating temperature 18-30°C (64-86°F)
Warm-up period ~20 minutes ~15 minutes
Max. # of marker templates practically unlimited
Max. # of markers concurrently tracked 100
Illumination range 50-100,000 Lux (20-400,000 Lux in HDR mode)
Max illumination variation between markers 6:13 (1000:1 in HDR mode)
Certifications IEC 60601-1, FCC Part 15 Class A, RoHS and WEEE compliant
Drift: temperature, pressure, light intensity, marker orientation, motion blur, exposure settings Insignificant
Hazard warnings issued Marker out of FOM; Thermal instability; Temperature out of range; Shadow over target
Operating system MS-Windows 7/Win 8/Win 10, Linux, 32bit and 64bit
Minimal computer hardware 2GHz CPU, 1 GB RAM, 1 GB disk space
 
  1. Single target (Xpoint), over 20,000 averaged positions at depths of 40-100cm
  2. Single target at a depth of 75cm, 150 Lux, 20ms shutter
  3. The range can be extended to 20:1 with black/gray targets under bright illumination

MicronTracker Fields of Measurement